The Labsphere DRA-CA-30I
Diffuse Reflectance accessory
For Cary 1, 3, 100 and 300 instruments
The Labsphere DRA-CA-30I is a 70 mm diameter DRA which fits inside
the sample compartment of a Cary 100 or Cary 300 instrument. Coated
with Spectralon (PTFE), this accessory offers flexibility when performing
both reflectance and transmittance measurements. The DRA CA-30I
is ideal for situations where an inexpensive, but high performance
integrating sphere, is required for samples which will fit inside
the sample compartment (approx. 150 mm diameter maximum). Measurements
of Color, Sun protection factor, liquids in cuvettes etc are all
possible with the DRA-CA-30I.
The DRA CA-30I is easily changed between transmittance and reflectance
measurements and allows the measurement of both Diffuse-only and
Total reflectance (a Spectralon reference plate is provided with
each accessory along with an 8° wedge to allow for Total reflectance
measurements).
The Labsphere CA-30I DRA does not need an Extended Sample Compartment
accessory for measuring reflectance of samples which have a thickness
of <25 mm. However, the extended
sample compartment will be needed for reflectance measurements
of samples which have a thickness >25 mm <45 mm.
Specifications
Wavelength range:
200 to 900 nm
Sphere diameter:
70 mm
Sphere coating:
Labsphere Spectralon
Sphere ports:
Reflectance port 21 mm Ø
Transmittance port 14 x 31 mm
Reference port 19.1 mm Ø
Detector port 35.7 x 19.4 mm ellipse
Transmittance sample holder: 13.3 x 9.5 mm
Port to sphere area
ratio for reflectance:
9.3%
Port to sphere ratio
with transmittance holder in place:
7.4%
Ordering Information
Description
Part number
Varian Labsphere DRA-CA-30I:
Includes spectralon reflectance disk (uncalibrated) and hard-copy
Operation Manual
0010063500
DRA-CA-30I Powder cell holder
set: Includes two blocks with a 1" D well in which powder
sample is placed. Quartz window covers sample so holder can
be positioned at port of sphere. One cell contains an uncalibrated
Spectralon standard to be used as a reference.
7910036600
10 mm pathlength cuvette holder
for transmittance port for DRA-CA-30I
7910028200
Fabric sample holder with UG-11
filter for DRA-CA-30I
7910027900
50 mm cuvette holder for transmission
port for DRA-CA-30I